• Vasileios Amoiridis's avatar
    iio: chemical: bme680: Fix sensor data read operation · 4241665e
    Vasileios Amoiridis authored
    A read operation is happening as follows:
    
    a) Set sensor to forced mode
    b) Sensor measures values and update data registers and sleeps again
    c) Read data registers
    
    In the current implementation the read operation happens immediately
    after the sensor is set to forced mode so the sensor does not have
    the time to update properly the registers. This leads to the following
    2 problems:
    
    1) The first ever value which is read by the register is always wrong
    2) Every read operation, puts the register into forced mode and reads
    the data that were calculated in the previous conversion.
    
    This behaviour was tested in 2 ways:
    
    1) The internal meas_status_0 register was read before and after every
    read operation in order to verify that the data were ready even before
    the register was set to forced mode and also to check that after the
    forced mode was set the new data were not yet ready.
    
    2) Physically changing the temperature and measuring the temperature
    
    This commit adds the waiting time in between the set of the forced mode
    and the read of the data. The function is taken from the Bosch BME68x
    Sensor API [1].
    
    [1]: https://github.com/boschsensortec/BME68x_SensorAPI/blob/v4.4.8/bme68x.c#L490
    
    Fixes: 1b3bd859 ("iio: chemical: Add support for Bosch BME680 sensor")
    Signed-off-by: default avatarVasileios Amoiridis <vassilisamir@gmail.com>
    Link: https://lore.kernel.org/r/20240606212313.207550-5-vassilisamir@gmail.com
    Cc: <Stable@vger.kernel.org>
    Signed-off-by: default avatarJonathan Cameron <Jonathan.Cameron@huawei.com>