Dave Penkler authored
Background: When performing a read on an instrument that is executing a function that runs longer than the USB timeout the instrument may hang and require a device reset to recover. The READ_STATUS_BYTE operation always returns even when the instrument is busy permitting to poll for the appropriate condition. This capability is referred to in instrument application notes on synchronizing acquisitions for other platforms. Signed-off-by: Dave Penkler <dpenkler@gmail.com> Signed-off-by: Greg Kroah-Hartman <gregkh@linuxfoundation.org>